[ Back]
ERBTEC RF amp shutdown Procedure
THE INTENT OF THESE TOP LEVEL MEASUREMENTS ARE TO GIVE YOU A QUICK GENERAL OVERVIEW OF THE FUNCTION OF THE RF SYSTEM. DO NOT ADJUST THE RF SYSTEM BASED ON THE VALUES YOU ATTAIN HERE. IF ADJUSTMENT IS REQUIRED, PERFORM THE ADJUSTMENTS WITH THE RELATED PROCEDURES MENTIONED BELOW.
| If a parameter is not mentioned here, leave the default value. |
| 1. ID: geservice | 8. SLICE THICK: 5mm |
| 2. COIL: HEAD or BODY (Depending on your requirements) | 9. TE: 20ms |
| 3. PHANTOM: DQA or Body Loader shell with phantom. | 10. TR: 300ms |
| 4. PSD: Spin-Echo | 11. FOV: Head=24cm Body=48cm |
| 5. MATRIX: 256x256 | 12. Select SCAN T/R |
| 6. WEIGHT: 300 lbs | 13. Increase TG to 200 (You must use Maximum TG) |
| 7. NEX: 1 | 14. Select [Scan T/R] when power Spectrum appears |
[ Next]