UP LS Actuator
Prerequisites
Procedure
- System Power must be turned OFF. Refer to Lockout / Tagout for System Cabinet PDU Main Breaker...
- Remove the scissor covers from the table. SCISSOR COVERS
- Loosen the defective Up LS actuator lock nut with the help of
7/16 spanner.
Figure 2. Loosen the Up LS actuator lock nut

- Remove the defective Up LS actuator Locknut and the Up LS Actuator.
Figure 3. Locknut and the Up LS Actuator

- Fix the new Up LS Actuator to the table top. And then the Lock nut to the Up LS actuator.
- Ensure the orientation of the Up Ls actuator such a way that
it actuates the limit switch when the table is taken to top most condition.
Figure 4. Up LS actuator

- Tighten the Lock nut of the Up LS Actuator.
Figure 5. Lock nut

- Fix the scissor covers from the table.
- Fix the FRP Covers of the Table.
1 Finalization
Procedure
- Turn On the System Power. Refer to Lockout / Tagout for System Cabinet PDU Main Breaker..
- Perform TABLE TOP HEIGHT ADJUSTMENT
- Check that table moves up and down smoothly.
- Check that Cradle moves in and out smoothly.