TABLE CONTROL BOARD CABLES

Prerequisites

Figure 1. DO NOT TAKE THE TABLE BEYOND THE MARKED LINE

Procedure

  1. System Power must be turned OFF. Refer to Lockout / Tagout for System Cabinet PDU Main Breaker...;
  2. Disconnect the Table from MR Magnet; take it out from the Magnet room. Refer to TAKE TABLE OUT FROM THE MAGNET ROOM
  3. Remove the scissor covers from the table. SCISSOR COVERS
  4. Remove the cable cramp and track and remove cables.

    Figure 2. Cable cramp and track

    1. Snap open cable clamp.

      Figure 3. Snap open cable clamp

    2. Snap open cable track at end link.

      Figure 4. Snap open cable track at end link

    3. Snap open cable track.

      Figure 5. Snap open cable track

  5. Remove the cables connected to the Up sensor Limit switch.

    Figure 6. Remove the cables

  6. Remove the Tabletop RH panel in the tabletop assembly.

    Figure 7. Remove the Tabletop RH panel

  7. Remove the cables connected to the Limit switch of the home sensor.

    Figure 8. Remove the cables

  8. Now the Control board cable assembly is free to remove from table.
  9. Now take the new Control Cable assembly.
  10. Rout the lengthier end of the cable in the tabletop.

    Figure 9. Rout

  11. Connect the cable lengthier end to the Limit switch of the home sensor.

    Figure 10. Connect the cable

  12. Fix the Tabletop RH panel in the tabletop assembly.

    Figure 11. Fix the Tabletop RH panel

  13. Connect the Shorter length end of the cable to the Limit switch.

    Figure 12. Connect the Shorter length end of the cable

  14. Rout the cables in the table by reverse order of removal.
  15. Fix the scissor covers to the table. Refer to SCISSOR COVERS.
  16. Move the Fixed Table and install to Dock Frame.
  17. Connect the Cables to the MR Gantry.