BACK       (Novice)                                                                            ERBTEC RF amp shutdown Procedure

Protocol Setup

THE INTENT OF THESE TOP LEVEL MEASUREMENTS ARE TO GIVE YOU A QUICK GENERAL OVERVIEW OF THE FUNCTION OF THE RF SYSTEM. DO NOT ADJUST THE RF SYSTEM BASED ON THE VALUES YOU ATTAIN HERE. IF ADJUSTMENT IS REQUIRED, PERFORM THE ADJUSTMENTS WITH THE RELATED PROCEDURES MENTIONED BELOW.

If a parameter is not mentioned here, leave the default value.
1. ID: geservice 8.  SLICE THICK: 5mm
2. COIL: HEAD or BODY (Depending on your requirements) 9.  TE: 20ms
3. PHANTOM: DQA or Body Loader shell with phantom. 10. TR: 300ms
4. PSD: Spin-Echo 11. FOV: Head=24cm Body=48cm
5. MATRIX: 256x256 12. Select SCAN T/R
6. WEIGHT: 300 lbs 13. Increase TG to 200 (You must use Maximum TG) 
7. NEX: 1 14. Select [Scan T/R] when power Spectrum appears

[ Next]