SRI Coil ID Stress Test

1 Diagnostic Link

Diagnostics >> System Function >> Patient Handling >> SRI Functional Tests

Diagnostics >> Hardware Location >> Magnet Room >> SRI Functional Tests

Figure 1. SRI Functional Tests Screen

2 Purpose

This tests the communication function of identifying coils. You may run this test on any coil plugged in to the system.

3 Components Tested

4 Requirements

The coil tested must have a coil ID.

5 Block Diagram

Figure 2. SRI Coil ID - Mega Switch Block Diagram

6 Test Sequence

The SCP commands the SRI to check for any coil connected. The SRI reads the coil presence information from the coil via the Mega Switch. If a coil is present, the SRI communicates with the coil through the Mega Switch.

7 Expected Results

If the coil is connected and is programmed with identification information, the coil id is read and identified on the diagnostic screen. The user can enter the number of iterations to be performed in the corresponding field. The test continues until all iterations have been performed or the user clicks Stop.

8 Etcetera

9 Links